• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. LOCKING SCREW INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. LOCKING SCREW INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 2901-90098
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that two everest locking screw inserter tips jammed intra-operatively.The drivers cross threaded with screw and tampered the threads.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report represents the second of the two inserters.
 
Event or Problem Description
It was reported that two everest locking screw inserter tips jammed intra-operatively.The drivers cross threaded with screw and tampered the threads.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report represents the second of the two inserters.
 
Additional Manufacturer Narrative
Visual inspection: upon review of the part, it was observed that the distal tip of the inner shaft showed signs of damage to the male hexalobe feature.Functional inspection: functional analysis revealed that the tip does not fully engage the hexalobe feature of a sample everest screw.Material analysis: the instrument was sent to imr materials analysis testing.No non-conformances were observed on the material during testing.It was suggested that these hexalobe tips may have been deformed due to incomplete insertion of the hexalobe tip on to the screw, or off-axis maneuvering of the instrument during surgical procedures.Device history records were reviewed for this lot, and no relevant manufacturing issues were identified.Complaint history records were reviewed for this lot and 4 similar complaints were identified.Per everest deformity surgical technique: after the pedicle pathway has been prepared and proper screw length and diameter have been determined, the appropriate implant is selected and loaded for screw insertion using the everest deformity screw inserter.The everest deformity screw inserter will work with both polyaxial and uni-planar everest screws.To load the everest deformity screw inserter, grasp the implant by the shaft of the screw and apply a downward force to engage the screw into the hexalobe fitting of the screw inserter shaft.Thread the silver knob in a clockwise direction until the implant is securely attached to the inserter.Pull the silver knob toward the handle to lock the inserter onto the screw.To disengage the screw inserter, pull down on the silver knob, gently turn in a counter-clockwise direction, and remove the inserter from the surgical field.Note: uniplanar screws are available upon request and may be selected if there is additional rotational deformity to be addressed (p.8).Ensuring that the screw is securely engaged in the inserter tip can assist in distributing forces evenly throughout the screw/inserter interface and reduce torque overloading.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
MDR Report Key10607326
Report Number3004774118-2020-00221
Device Sequence Number3830845
Product Code LXH
UDI-Device Identifier10888857040564
UDI-Public10888857040564
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2901-90098
Device Lot NumberGUYU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Initial Date Received by Manufacturer 08/31/2020
Supplement Date Received by Manufacturer12/30/2020
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date01/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-