| Model Number |
FRED4508-PMA |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformance associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that upon deployment of the fred in the internal carotid artery, the device was partially expanded.The stent was completely expanded with additional guide wire manipulations.There was no reported patient injury.The patient is currently reported to be doing well.
|
| |
|