MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Medical Device Problem Code Defibrillation/Stimulation Problem (1573)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product ,and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during biomed testing, the device displayed "defib fault 72," "defib fault 76", and "defib disabled" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation; the customer's report was observed during review of the device activity log.However, the device was put through extensive testing power cycling, bench handling, and defib functionality stress testing without duplicating the report.The pace/defib engine board was replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key10607850
Report Number1220908-2020-03059
Device Sequence Number3985652
Product Code MKJ
UDI-Device Identifier00847946001988
UDI-Public00847946001988
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K060559/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/10/2020
Supplement Date Received by Manufacturer09/10/2020
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-