| Model Number |
BXA057902A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Renal Failure (2041); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Concomitant medical devices: as it is not known which device if any contributed to the acute kideney injury, the reported device and all of the following were investigated: bxa067902a sn#: (b)(4), udi: (b)(4), bxa057902a sn#: (b)(4) udi: (b)(4), bxa067902a sn#: (b)(4), udi: (b)(4), bxa067902a sn#: (b)(4), udi: (b)(4).(b)(4).A review of the manufacturing records for the devices verified the lots met all pre-release specifications.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2020 a patient underwent endovascular aortic debranching with investigational tambe devices.Gore® viabahn® vbx balloon expandable endoprostheses were utilized in both renal arteries.On (b)(6) 2020 the patient exhibited acute kidney injury, which was reported to have been possibly device related.The kidney injury was not specified to a certain side, but rather detected as a general finding in a kidney function test.No treatment was rendered and the case is ongoing.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2020 a patient underwent endovascular aortic debranching with investigational tambe devices.Gore® viabahn® vbx balloon expandable endoprostheses were utilized in both renal arteries.On (b)(6) 2020 the patient exhibited acute kidney injury, which was reported to have been possibly device related.The kidney injury was not specified to a certain side, but rather detected as a general finding in a kidney function test.No treatment was rendered and the case is ongoing.Upon further investigation, the acute kidney injury was not found to be due to any device-related issue.There was no occlusion or stenosis.The renal arteries were patent.
|
| |
|
Additional Manufacturer Narrative
|
|
B.5.Updated.Medwatch # 2017233-2020-01311 was sent in error.Additional received information determined that this event is not reportable to the fda and therefore the medwatch (and supplementals) will be retracted.
|
| |
|