• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING SL; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING SL; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24703
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/22/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left popliteal artery.After ivus was performed, a 4.0mm x 150mm x 150cm sterling balloon catheter was advanced for dilation.However, during inflation at 4 atmospheres the balloon ruptured and would not inflate further.The device was removed and the patient condition was stable and unharmed.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling sl balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast and blood in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a longitudinal tear 1mm long starting 53mm from the distal marker band.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported balloon ruptured.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left popliteal artery.A 4.0mm x 150mm x 150cm sterling balloon catheter was selected for use.The balloon was tracked appropriately over the wire to target in the popliteal artery.The balloon was inflated at 4 atmospheres when it burst and would not inflate further.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING SL
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10608258
Report Number2134265-2020-13548
Device Sequence Number16844783
Product Code DQY
UDI-Device Identifier08714729782384
UDI-Public08714729782384
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K093720
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/17/2022
Device Model Number24703
Device Catalogue Number24703
Device Lot Number0025729980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer11/12/2020
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date11/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-