| Model Number |
24703 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the left popliteal artery.After ivus was performed, a 4.0mm x 150mm x 150cm sterling balloon catheter was advanced for dilation.However, during inflation at 4 atmospheres the balloon ruptured and would not inflate further.The device was removed and the patient condition was stable and unharmed.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a sterling sl balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast and blood in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a longitudinal tear 1mm long starting 53mm from the distal marker band.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported balloon ruptured.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the left popliteal artery.A 4.0mm x 150mm x 150cm sterling balloon catheter was selected for use.The balloon was tracked appropriately over the wire to target in the popliteal artery.The balloon was inflated at 4 atmospheres when it burst and would not inflate further.
|
| |
|
Search Alerts/Recalls
|