| Model Number |
VITAL SIGNS¿ ORAL AIRWAY |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.Therefore, no root cause could be determined yet.
|
| |
|
Event or Problem Description
|
|
The customer reported that the small, yellow piece of plastic on 3590 vital signs¿ oral airway slid out of the oral airway into patient's pharynx during intubation.The customer confirmed that it was removed with hagil forceps prior to case start preventing harm to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: vyaire medical was able to determined the reported issue.One opened decontaminated sample was received for evaluation.It was observe that part number r3090h2 has excess loose yellow material.The root cause was traced to the material - material does not meet manufacturers specifications.As a corrective and preventive action, existing scar (b)(4) was initiated.
|
| |
|