BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10607
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A synergy ii us mr 4.00 x 38mm stent delivery system was returned for analysis.An examination of the crimped stent identified proximal stent damage with stent strut lifted and bunched distally.The undamaged crimped stent outer diameter was within max crimped stent profile measurement.The balloon was reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found multiple hypotube kinks.A examination of the shaft polymer extrusion found no issues.An examination of the tip found no issues.No other issues were identified during the product analysis.
 
Event or Problem Description
Reportable based on device analysis completed on 09sep2020.It was reported that crossing difficulties were encountered.A 4.00 x 38mm synergy drug eluting stent was advanced for treatment; however, the stent could not cross the lesion.There were no patient complications reported.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10609752
Report Number2134265-2020-13563
Device Sequence Number2802876
Product Code NIQ
UDI-Device Identifier08714729840435
UDI-Public08714729840435
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2022
Device Model Number10607
Device Catalogue Number10607
Device Lot Number0024998257
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/09/2020
Initial Report FDA Received Date09/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-