| Model Number |
10492730 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
09/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer stated that the operator was wearing gloves.The sample port is not available to be returned for investigation.The operator stated that the sample port appeared to be clean.The system is designed so that there is no carry-over between each sample through calibration wash cycles which helps minimize exposure.The system washes after every sample to aid in keeping the system clean and decrease and minimize any biohazard exposure.The customer stated the operator is doing well.No further information or details provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that the operator states that he tried to replace a measurement cartridge in the rp 500 and noticed a door error and electronic error.He then tried to push the door shut and noticed his right hand, index finger, punctured the sample port needle.He immediately bled out the cut and followed occupational health procedures and policies and went to the er to get treated.
|
| |
|