• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MODIFIED JII TIBIA CUTTING BLOCK RIGHT; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MODIFIED JII TIBIA CUTTING BLOCK RIGHT; TEMPLATE Back to Search Results
Catalog Number E0015575-2
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery (instrument inside the patient) the trocar pin got bound up in the pin hole.Trocar pin broke off in pin hole.The procedure was completed without delay with the same device.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that, this complaint from the united states reports that a trocar pin broke off in the pin hole.Reportedly, the procedure was completed without delay with the same device.Smith and nephew has not received adequate materials (operative reports or x-rays) to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.The impact to the patient cannot be determined because it was not reported, whether the broken pin was retained.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Damage from misuse or rough handling are likely probable causes of the reported event.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODIFIED JII TIBIA CUTTING BLOCK RIGHT
Common Device Name
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10609921
Report Number1020279-2020-04989
Device Sequence Number2410091
Product Code HWT
UDI-Device Identifier00885556701645
UDI-Public885556701645
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberE0015575-2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/02/2020
Supplement Date Received by Manufacturer08/05/2021
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date08/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-