Model Number 00884450009482
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that post-drainage catheter placement within the patient's biliary system, the drainage catheter became damaged at the bend 7 days later and would not continue to drain.The drainage catheter was replaced for this patient.
 
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.The device was examined visually.The complaint could not be confirmed.The root cause is attributed to off label use, the device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLVE LOCKING DRAINAGE CATHETER
Common Device Name
DRAINAGE CATHETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
14646 kirby drive
houston TX 77047
MDR Report Key10609993
Report Number3010665433-2020-00037
Device Sequence Number1192767
Product Code GBX
UDI-Device Identifier00884450009482
UDI-Public00884450009482
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model Number00884450009482
Device Catalogue NumberRLC-8-038/CNB
Device Lot NumberE1620806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Initial Date Received by Manufacturer 09/13/2020
Supplement Date Received by Manufacturer11/11/2020
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date11/18/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-