Model Number LTF-S190-5
Device Problems No Display/Image (1183); Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/05/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.In the evaluation of omsc the following was confirmed; the reported phenomenon was not reproduced.The subject device was disassembled and investigated, but there was no problem inside the subject device such as water ingress, failure of soldering and wiring.However when the electrical circuit board with the condition disassembled the subject device was heated, the endoscopic image disappeared occasionally.Therefore omsc identified that this phenomenon was attributed to the electrical component failure of the electrical circuit board.The exact cause of the electrical component failure could not be conclusively determined, however there was the possibility the cause of the electrical component failure might be the deterioration, overcurrent or static electricity.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during preparation for therapeutic procedure with the subject device, the communication error occurred and the endoscopic image was not displayed.The procedure was completed using another device.There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10610876
MDR Text Key223661984
Report Number8010047-2020-07032
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080948
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-