BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 10620
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/14/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: synergy ous mr 2.50 x 24mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage.Stent struts from the distal end of the stent were lifted and pulled distally.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 09sep2020 it was reported that crossing difficulties were encountered.The target lesion was located in the moderately tortuous and severely calcified distal left anterior descending artery.A 2.50x24mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications nor injuries were reported.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10611000
MDR Text Key209351541
Report Number2134265-2020-13577
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/08/2020
Device Model Number10620
Device Catalogue Number10620
Device Lot Number0023829413
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
-
-