MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1780KPK 670G PATHWAY BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
|
Back to Search Results |
|
Model Number MMT-1780KPK |
Device Problems
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/22/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Information received by medtronic indicated that the retainer ring was chipped and the reservoir was not able to lock in place.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Device was received with a broken reservoir tube lip, partially broken retainer and missing reservoir tube o-ring.Device p-cap / test reservoir does not lock in place properly and unable to perform the displacement test due to the partially broken retainer and broken reservoir tube lip.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0956-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|