• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN IDRIVE ULTRA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN IDRIVE ULTRA; STAPLE, IMPLANTABLE Back to Search Results
Model Number IDRVULTRA1
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during servicing, after normal high temperature and high-pressure disinfection, the adapter loading ring on the handle fell off.The product was back and not used.There was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the handle's quick connect assembly was disengaged.Scratches were noted on the inner wall of the collar, indicating that the collar and ball ramp were pressed together.The ball ramp was inspected and no visual deformation or imperfections were seen.It was reported that a component disengaged from the device.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.This issue can occur if there was an out of specification part in combination with exposure to heat.This will cause a defect outside of the operating room during autoclaving, with no proximity to the patient.The condition will not cause any patient harm.When the collar is still in position on the ring, it cannot separate under normal operating conditions.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDRIVE ULTRA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key10611796
Report Number1219930-2020-04054
Device Sequence Number13546284
Product Code GDW
UDI-Device Identifier10884521142794
UDI-Public10884521142794
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121510
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIDRVULTRA1
Device Catalogue NumberIDRVULTRA1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2020
Initial Date Received by Manufacturer 09/15/2020
Supplement Date Received by Manufacturer11/13/2020
12/04/2020
Initial Report FDA Received Date10/01/2020
Supplement Report FDA Received Date12/04/2020
12/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-