BIOTRONIK SE & CO. KG PROTEGO T 65; ICD LEAD Back to Search Results
Model Number 379967
Medical Device Problem Codes Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 09/07/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approx.66 months after the implantation the patient received inappropriate shocks due to ventricular oversensing.The lead was capped and replaced.Apart from the shocks no further patient side effects were reported.
 
Additional Manufacturer Narrative
As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The analysis is thus based on the inspection of the manufacturing documents of the device as well as on the provided data.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.The analysis of the provided trend data and iegm recordings, acquired between (b)(6) 2017 and (b)(6) 2018, gained no further information.In spite of the available information, no conclusion can be drawn regarding the root cause of the clinical observation.An analysis of the lead itself would be necessary to determine the root cause.Should additional information or the device itself become available for analysis, the investigation will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGO T 65
Common Device Name
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key10612318
Report Number1028232-2020-04190
Device Sequence Number3004232
Product Code NVY
UDI-Device Identifier04035479121686
UDI-Public04035479121686
Combination Product (Y/N)N
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2016
Device Model Number379967
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/30/2020
Supplement Date Received by Manufacturer04/14/2021
Initial Report FDA Received Date10/01/2020
Supplement Report FDA Received Date04/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-