Model Number 319.006
Device Problem Material Twisted/Bent (2981)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Part number:319.006, synthes lot number: h663667, supplier lot number: n/a, release to warehouse date: 17apr2019, expiration date: n/a, supplier: (b)(4).No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Visual inspection: the depth gauge for 2.0mm and 2.4mm screws (part # 319.006 / lot # h663677) was received at us customer quality (cq).The needle of the device was observed to be bent.The needle is also observed to be loose from the slider as the threads connecting to the slider could be observed to be unscrewed.No other defects were identified.The received condition was consistent with the complaint condition thus the complaint was confirmed.Device failure/defect identified? yes; the device needle was bent.Dimensional inspection: needle shaft diameter=conforming.Document/specification review: drawing(s) reviewed: (current & manufactured revisions).Conclusion: the overall complaint was confirmed for the received depth gauge for 2.0mm and 2.4mm screws as the needle was bent.Although no definitive root-cause can be determined it is possible the device experienced unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.
 
Event Description
It was reported that on an unknown date during a routine incoming inspection of the loaner set, it was observed that the depth gauge was bent.There was no patient involvement.This report is for one (1) depth gauge for 2.0 mm and 2.4 mm screws.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key10613045
MDR Text Key209404550
Report Number2939274-2020-04408
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982189943
UDI-Public10886982189943
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2020
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number319.006
Device Catalogue Number319.006
Device Lot NumberH663667
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-