| Model Number |
M00568530 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an endovive one step button was used during a percutaneous endoscopic gastrostomy (peg) placement procedure on (b)(6) 2020.According to the complainant, during the procedure, the internal bolster detached inside the patient's stomach.Reportedly, the detached portion was removed.There were no reported patient complications as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6 (device codes: problem code 2907 captures the reportable event of internal bolster detached.Block h10: a portion of the one step button was returned.Visual examination of the returned device found that the c-flex tube broken.The button returned with one part of the c-flex tube connected, another part of the c-flex tube stretched and waste from the procedure were presented on the device.Microscope inspection revealed that the c-flex tube is narrowed near the broken area indicating stretching of material and was irregular.With the available information, boston scientific concludes that the c-flex tube broken was inspected under magnification and it showed an irregular shape and stretched, indicating that it was submitted to some tension under procedure.Based on the condition of the returned device, engineers determined that this may be related to some technique applied during procedure.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on event which led to the reported event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an endovive one step button was used during a percutaneous endoscopic gastrostomy (peg) placement procedure on (b)(6)2020.According to the complainant, during the procedure, the internal bolster detached inside the patient's stomach.Reportedly, the detached portion was removed.There were no reported patient complications as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
|
| |
|