• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 48MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 48MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-01-048
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon implanted a pinnacle 100 cup and then wanted to insert the hole eliminator screw.This could just be screwed through and showed no resistance/hold/stop.When surgeon noticed that the screw was probably not running tight, he couldn't turn it back either.The locking screw is clearly too deep.No surgery delay reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Review of the attached x-ray image confirmed the reported complaint.A search of the depuy nonconformance (nc) quality system found no manufacturing related nc¿s associated with this product/lot combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: dhr has been reviewed and no deviations or anomalies were found.A search of the depuy nonconformance (nc) quality system found no manufacturing related nc¿s associated with this product/lot combination.
 
Event or Problem Description
Additional information received indicated that the screw was loose and could not be turned back because it slipped thru.The event happened during a primary surgery.Affected side was left.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 100 ACET CUP 48MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10613229
Report Number1818910-2020-21348
Device Sequence Number10877146
Product Code LPH
UDI-Device Identifier10603295008538
UDI-Public10603295008538
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K000306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-01-048
Device Catalogue Number121701048
Device Lot Number9574213
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/18/2020
Supplement Date Received by Manufacturer10/07/2020
12/15/2020
Initial Report FDA Received Date10/01/2020
Supplement Report FDA Received Date10/19/2020
12/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; PINN MAR +4 10D 28IDX48OD; APEX HOLE ELIM POSITIVE STOP; PINN MAR +4 10D 28IDX48OD
Patient Age18 YR
-
-