• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ASPIRATION NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ASPIRATION NEEDLE Back to Search Results
Model Number NA-201SX-4022
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to olympus for evaluation.The cause of the reported event cannot be determined.If additional information becomes available at a later time, this report will be supplemented.
 
Event or Problem Description
It was reported that during an endobronchial ultrasound (ebus), while completing the first sampling puncture, the needle broke off and was stuck in the patient.As a result, the fragment was retrieved from the patient.The procedure was completed as intended.No replacement devices were used.
 
Additional Manufacturer Narrative
This report is being supplemented to provide the findings of the investigation.New information is reported in h6, and h10.The device referenced in this report was not returned to olympus for physical evaluation, it was discarded by the customer along with the packaging, so the lot number of the complaint device could not be provided.Since the production lot number of the device was unknown, olympus checked the device history records (dhr )for the past year (october 2019 to september 2020) from the date of occurrence, and found that there were no abnormalities in the items related to the reported event.The instructions for use shipped with the device provides the customer with the following instructions related to the the event: ·when inserting the instrument into the endoscope, the distal end of the needle tube may be bent.When piercing the target, confirm the distal end of the sheath and needle tube in the endoscopic field of view and/or ultrasound image while considering such bending.Otherwise, patient injury such as perforation, bleeding, or mucous membrane damage may occur.Do not try to straighten a bent or deformed needle with your hands because the needle may break.Use a spare needle instead.·do not push the needle slider suddenly.Doing so may cause sudden needle protrusion, resulting in perforation, bleeding, or mucous membrane damage.It can also damage the instrument.Conclusion/summary: since the actual product could not be confirmed and there were no abnormalities in the manufacturing records, the cause of the occurrence could not be identified.A past similar case indicates that a buckled needle tube breaks if straightening force is applied to the buckling.It is also known the needle weakens when repeatedly bent and straightened as the metal used for it has such a nature.The needle tube likely broke off by the mechanism below.1.The needle tube buckled significantly due to force to bend it, which was possibly generated by movements of the patient during the procedure.2.The buckled needle tube then received force to straighten it.3.The bending and straightening reduced the strength of the needle tube until and finally broke it off.However, the exact cause of the needle break could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ASPIRATION NEEDLE
Common Device Name
SINGLE USE ASPIRATION NEEDLE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10613559
Report Number8010047-2020-07080
Device Sequence Number19477492
Product Code FCG
UDI-Device Identifier04953170389900
UDI-Public04953170389900
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K050503
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNA-201SX-4022
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/08/2020
Supplement Date Received by Manufacturer11/11/2020
Initial Report FDA Received Date10/01/2020
Supplement Report FDA Received Date12/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-