Combination product: yes the returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.In addition, the angiographic material provided was reviewed.The technical investigation of the returned device revealed that the hypotube has fractured about 14cm distal to the distal end of the kink protector.Several small kinks were observed in the hypotube.The cross section of the hypotube is no longer circular but compressed and the polymer coating is plastically deformed.The findings indicate that the hypotube was most probably significantly bent outside of the patients body and eventually fractured during the lesion crossing attempt.Moreover, the stent is displaced by 7.5mm distally, but the balloon is still well folded and shows no signs of inflation.Stent imprints on the exposed balloon surface indicate that the stent was initially crimped in between the two radiopaque markers.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in process and final inspections.During final inspection and inner packaging every device is visually inspected for kinks.As a part of final inspection every stent system also undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.The stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|