• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938740
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 09/08/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 2000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was defective.The defect was further described as a leak coming from the middle port.This issue was identified during setup and preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: device manufactured between november 06, 2019 to november 07, 2019.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional test was performed which revealed a leak at the spike port bonding area between the spike port tubing and the spike port connector.The reported condition was verified.The cause was not determined, however, a likely cause of inadequate or lack of cyclohexanone being applied to the cap tubing when it was inserted to the spike port during the manufacturing process was noted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10613765
MDR Text Key209398682
Report Number1416980-2020-06119
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477244
UDI-Public(01)00085412477244
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2022
Device Catalogue NumberH938740
Device Lot Number60213295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2020
Date Manufacturer Received10/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-