| Model Number |
3001-06575 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A physician reported that a patient underwent revision surgery to replace a fractured mesa polyaxial screw.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a mesa small stature deformity polyaxial screw broke post-operatively at the sacroiliac level.The screw was returned, visually and microscopically inspected.Upon review of the part, it was observed that the screw head assembly was separated from the screw shank.The proximal ball feature of the screw shank exhibited deformations suggesting that the screw was positioned at an angle.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.A review of complaint history associated with the subject catalog number was performed and no adverse trends were observed.The tulip disengagement was also confirmed via x-ray.It is possible that non-union contributed to the failure.According to the company representative in the field, it could not be determined if fusion was achieved.Non-union could allow for continued dynamic motion within the construct, which may contribute to a failure of this nature.Post-operative patient activity may have also contributed to the failure.
|
| |
|
Event or Problem Description
|
|
A physician reported that a patient underwent revision surgery to replace a fractured mesa polyaxial screw.
|
| |
|