| Model Number |
6260-9-340 |
| Medical Device Problem Codes |
Material Erosion (1214); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Ambulation or Postural Difficulties (2544)
|
| Date of Event |
09/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
|
| |
|
Event or Problem Description
|
|
As reported: "patient came to office on crutches.Revised femoral stem.Accolade primary stem neck failure, trunnion.Trunnion wore away.Femoral head popped off.Revision surgery to restoration modular.".
|
| |
|
Additional Manufacturer Narrative
|
|
The event was confirmed via photos and medical review.Reported event: an event regarding disassociation and fretting involving a lfit v40 cocr head that was mated with an accolade stem.The event was confirmed via photos and medical review.Method & results: product evaluation and results: the reported device was not returned however photographs were provided for review.The photographs note the following: metal head was covered in biological material and black material was observed on the taper and the accolade stem trunnion was severely worn confirming the reported disassociation and fretting events.Medical records received and evaluation: a review of the provided medical information by a clinical consultant indicated: confirmation: the event can be confirmed.Dissociation of head from stem with worn trunnion.Hazard/root cause: a hazard exists.The v40 cocr lfit head dissociated from the stem.There existed severe trunnion wear.The event is likely related to the lfit recall.Obtaining and examining the device may confirm the finding.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: lot specific voluntary recall v40 - recall -2249697-08/29/2016-007r was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
|
| |
|
Event or Problem Description
|
|
As reported: "patient came to office on crutches.Revised femoral stem.Accolade primary stem neck failure, trunnion.Trunnion wore away.Femoral head popped off.Revision surgery to restoration modular.".
|
| |
|