Model Number 6260-9-340
Medical Device Problem Codes Material Erosion (1214); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Ambulation or Postural Difficulties (2544)
Date of Event 09/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Event or Problem Description
As reported: "patient came to office on crutches.Revised femoral stem.Accolade primary stem neck failure, trunnion.Trunnion wore away.Femoral head popped off.Revision surgery to restoration modular.".
 
Additional Manufacturer Narrative
The event was confirmed via photos and medical review.Reported event: an event regarding disassociation and fretting involving a lfit v40 cocr head that was mated with an accolade stem.The event was confirmed via photos and medical review.Method & results: product evaluation and results: the reported device was not returned however photographs were provided for review.The photographs note the following: metal head was covered in biological material and black material was observed on the taper and the accolade stem trunnion was severely worn confirming the reported disassociation and fretting events.Medical records received and evaluation: a review of the provided medical information by a clinical consultant indicated: confirmation: the event can be confirmed.Dissociation of head from stem with worn trunnion.Hazard/root cause: a hazard exists.The v40 cocr lfit head dissociated from the stem.There existed severe trunnion wear.The event is likely related to the lfit recall.Obtaining and examining the device may confirm the finding.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: lot specific voluntary recall v40 - recall -2249697-08/29/2016-007r was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Event or Problem Description
As reported: "patient came to office on crutches.Revised femoral stem.Accolade primary stem neck failure, trunnion.Trunnion wore away.Femoral head popped off.Revision surgery to restoration modular.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 40MM/+8
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10617165
Report Number0002249697-2020-02030
Device Sequence Number16236389
Product Code LPH
UDI-Device Identifier07613327032765
UDI-Public07613327032765
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K061434
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2015
Device Model Number6260-9-340
Device Catalogue Number6260-9-340
Device Lot NumberMJA7XT
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/08/2020
Supplement Date Received by Manufacturer11/04/2020
Initial Report FDA Received Date10/01/2020
Supplement Report FDA Received Date11/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-08/29/2016-007R
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient Weight147
-
-