| Model Number |
3245 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
07/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
¿date of event¿ is estimated.The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that patient experienced ineffective therapy.Diagnostic testing revealed high impedance on multiple contacts of the lead.Reprogramming did not resolve the issue.Surgery may occur to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h6 - method code has been updated to 4114.
|
| |
|
Event or Problem Description
|
|
Additional information was received that surgery occurred during which the lead was found to be fractured.During the surgery, the lead was explanted and replaced to address the issue.
|
| |
|