Model Number 3245
Medical Device Problem Codes Fracture (1260); High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 07/25/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
¿date of event¿ is estimated.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
It was reported that patient experienced ineffective therapy.Diagnostic testing revealed high impedance on multiple contacts of the lead.Reprogramming did not resolve the issue.Surgery may occur to address the issue.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Additional Manufacturer Narrative
Additional information: h6 - method code has been updated to 4114.
 
Event or Problem Description
Additional information was received that surgery occurred during which the lead was found to be fractured.During the surgery, the lead was explanted and replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE 44C LEAD KIT, 60CM LENGTH
Common Device Name
SCS PADDLE LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key10617768
Report Number1627487-2020-31852
Device Sequence Number14063206
Product Code LGW
UDI-Device Identifier05414734402088
UDI-Public05414734402088
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2013
Device Model Number3245
Device Catalogue Number3245
Device Lot Number3463521
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/10/2020
Supplement Date Received by Manufacturer10/07/2020
11/03/2020
Initial Report FDA Received Date10/01/2020
Supplement Report FDA Received Date10/13/2020
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Weight57
-
-