• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. V SPLITTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. V SPLITTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71367561
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/03/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection, plastic tip was found to be broken.No patient involved.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirms the device fractured into two pieces.Both pieces were returned.The device was manufactured in 2001.The device exhibits signs of extreme wear and use.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V SPLITTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10618379
Report Number1020279-2020-05065
Device Sequence Number14135631
Product Code LXH
UDI-Device Identifier03596010403131
UDI-Public03596010403131
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2011
Device Model Number71367561
Device Catalogue Number71367561
Device Lot Number01BM09312
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Initial Date Received by Manufacturer 09/03/2020
Supplement Date Received by Manufacturer11/30/2020
Initial Report FDA Received Date10/02/2020
Supplement Report FDA Received Date12/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-