• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. LEAD ADAPTOR KIT; ADAPTOR, LEAD, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. LEAD ADAPTOR KIT; ADAPTOR, LEAD, PACEMAKER Back to Search Results
Model Number 5866-38M
Medical Device Problem Codes Break (1069); Material Integrity Problem (2978)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2009
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: 4968-25 lead, implant date: (b)(6) 2004; a2dr01 ipg, implant date: (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while dissecting the pocket high undefined impedance and no capture at high output were noted on the right ventricular (rv) lead.Insulation degradation and tissue ingrowth were also noted on the y adapter.The adapter was explanted and the lead was partially removed.The implantable pulse generator (ipg) was explanted and replaced for an unknown reason.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while dissecting the pocket high undefined impedance and no capture at high output were noted on the right ventricular (rv) lead.Insulation degradation and tissue ingrowth were also noted on the y adapter.The adapter was explanted and the lead was partially removed.The implantable pulse generator (ipg) was explanted and replaced for an unknown reason.No patient complications have been reported as a result of this event.It was also reported that the right atrial (ra) lead was fractured and that one rv lead was fractured and that both rv lead exhibited high thresholds.The leads remain in use.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was also reported that the right atrial (ra) lead was fractured and that one rv lead was fractured and that the other rv lead exhibited high thresholds.The leads remain in use.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned, analyzed, and the outer insulation of the lead was extrinsically breached due to a cut.The proximal conductor of the lead became extrinsically distorted due to kinking/buckling.The distal conductor was extrinsically distorted due to kinking/buckling.Visual analysis of the lead indicated apparent explant damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
It was further reported that the device was explanted as it was nearing elective replacement indicator (eri).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD ADAPTOR KIT
Common Device Name
ADAPTOR, LEAD, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key10618647
Report Number2182208-2020-02200
Device Sequence Number3620188
Product Code DTD
UDI-Device Identifier00681490136907
UDI-Public00681490136907
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K911302
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Device Explanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/28/2005
Device Model Number5866-38M
Device Catalogue Number5866-38M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2020
Initial Date Received by Manufacturer 09/09/2020
Supplement Date Received by Manufacturer10/07/2020
10/12/2020
11/06/2020
11/24/2020
Initial Report FDA Received Date10/02/2020
Supplement Report FDA Received Date10/09/2020
11/03/2020
11/09/2020
12/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5071-35 LEAD, 5071-35 LEAD.; 5071-35 LEAD, 5071-35 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age23 YR
Patient Weight72
-
-