B. BRAUN SURGICAL, S.A. DAFILON BLUE 5/0 (1) 45CM DS19; OTHER SUTURE Back to Search Results
Model Number C0932191
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available a follow up report will be submitted.
 
Event Description
It was reported that there was an issue with dafilon suture.The client reported that the needle was detached from the thread when pulling out the suture from package.The defect was found prior to use, there is no patient involvement.
 
Manufacturer Narrative
Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) of this code-batch.There are no units in our stock.We have received two open and unused samples.The needle is attached to the thread and the suture is wound on the pack.We have tested the needle attachment strength of the samples received and the results fulfil the requirements of the european pharmacopoeia (ep): 0.54 kgf in average and 0.52 kgf in minimum (ep requirements: 0.23 kgf in average and 0.11 kgf in minimum) reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Needle attachment strength results before releasing the product were 0.51 kgf in average and 0.38 kgf in minimum and fulfilled ep requirements.Final conclusion: although the results of the samples received fulfil the specifications of european pharmacopoeia/ b.Braun surgical specifications, we take note of this incidence in order to assess if new or additional actions are needed.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 5/0 (1) 45CM DS19
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa, 121
rubí, 08191
SP  08191
MDR Report Key10618693
MDR Text Key209587640
Report Number3003639970-2020-00357
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
PMA/PMN Number
K151165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0932191
Device Catalogue NumberC0932191
Device Lot Number620116
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Date Manufacturer Received10/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-