Model Number 0684-00-0576-01
Device Problems Difficult to Insert (1316); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/11/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that during insertion of the intra-aortic-balloon (iab), the iab membrane became unwrapped and could not fit through the sheath.A new iab was opened immediately, and inserted with no issues.There was no patient injury, or adverse event reported.
 
Event Description
It was reported that during insertion of the intra-aortic-balloon (iab), the iab membrane became unwrapped and could not fit through the sheath.A new iab was opened immediately and inserted with no issues.There was no patient injury or adverse event reported.
 
Manufacturer Narrative
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter with the one-way valve attached.The sheath was not returned for evaluation.The one-way valve was vacuum tested and it held vacuum.A laboratory insertion test was unable to be performed due to the membrane being unfurled.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.We are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.The device meets all product specifications.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 8FR. 50CC
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key10619029
MDR Text Key209561857
Report Number2248146-2020-00519
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108605
UDI-Public10607567108605
Combination Product (y/n)N
PMA/PMN Number
K112327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/23/2023
Device Model Number0684-00-0576-01
Device Catalogue Number0684-00-0575
Device Lot Number3000117327
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Date Manufacturer Received10/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age78 YR
Patient Weight95
-
-