Model Number 0684-00-0576-01 |
Device Problems
Difficult to Insert (1316); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that during insertion of the intra-aortic-balloon (iab), the iab membrane became unwrapped and could not fit through the sheath.A new iab was opened immediately, and inserted with no issues.There was no patient injury, or adverse event reported.
|
|
Event Description
|
It was reported that during insertion of the intra-aortic-balloon (iab), the iab membrane became unwrapped and could not fit through the sheath.A new iab was opened immediately and inserted with no issues.There was no patient injury or adverse event reported.
|
|
Manufacturer Narrative
|
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter with the one-way valve attached.The sheath was not returned for evaluation.The one-way valve was vacuum tested and it held vacuum.A laboratory insertion test was unable to be performed due to the membrane being unfurled.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.We are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.The device meets all product specifications.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint# (b)(4).
|
|