SMITH & NEPHEW, INC. FEMORAL PEG DRILL; BIT, DRILL Back to Search Results
Model Number 74016421
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/04/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported an increased resistance while drilling with the femoral peg drill during a tka surgery, also wear patterns appear to be forming on the shafts of the drill bit; there is no information regarding how the procedure was completed nor if there was any delay.No other patient injury or other complications were reported.
 
Additional Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is worn with nicks, rendering the device inoperable.The device was manufactured in 2018 and shows signs of extensive use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL PEG DRILL
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10620757
Report Number1020279-2020-05095
Device Sequence Number2793136
Product Code HTW
UDI-Device Identifier00885556692233
UDI-Public00885556692233
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74016421
Device Catalogue Number74016421
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/04/2020
Supplement Date Received by Manufacturer02/01/2021
Initial Report FDA Received Date10/02/2020
Supplement Report FDA Received Date02/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-