• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE POST OP 10X35CM CTN 20; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE POST OP 10X35CM CTN 20; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66000716
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/08/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that opsite post op size 35x10 has adhesion problem and does not properly stick to the site.No harm or injury reported.
 
Additional Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.A complaints history review was carried out using the lot and part numbers provided, there have been no further complaints reported with this failure mode in the past three years.The device was used for treatment.It was reported that the dressing did not adhere to the patient¿s skin satisfactorily.As no samples were returned a thorough product evaluation could not be carried out.Low adhesion may be caused by adhesion levels of the raw material used to make the film of this dressing.There are various controls in place during the manufacturing process of both the raw material and the dressing to identify when the adhesion is not in acceptable range, as per the product specification.There are also regular tests carried out to monitor the finished product.As with all adhesive products, the skin site must be thoroughly cleaned and dried prior to application.If there is any moisture on the skin surrounding the catheter, the adhesive performance of this dressing may be compromised.If the dressing becomes wet whilst applied, it may have to be removed and a new dressing applied.We have not been able to confirm a relationship between the event and the device.The root cause remains undetermined.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE POST OP 10X35CM CTN 20
Common Device Name
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10621300
Report Number8043484-2020-03095
Device Sequence Number11514261
Product Code KGX
UDI-Device Identifier05000223441456
UDI-Public05000223441456
Combination Product (Y/N)N
Initial Reporter CountryIN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number66000716
Device Lot Number1712
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/08/2020
Supplement Date Received by Manufacturer10/21/2020
Initial Report FDA Received Date10/02/2020
Supplement Report FDA Received Date10/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-