Model Number 97715
Device Problems Energy Output Problem (1431); Inappropriate/Inadequate Shock/Stimulation (1574); Environmental Compatibility Problem (2929)
Patient Problems Device Overstimulation of Tissue (1991); Electric Shock (2554)
Event Date 09/29/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Patient reported yesterday (b)(6) 2020: she put on her pants that the material was leopard skin metallic (for a halloween costume) and she instantly was getting "shocky" sensations from stim so she turned stim down and finally turned it off.She described it as "hyperstimulation".She stated she came home and took the pants off but since then she is still feeling that sensation when she turns stimulation on.It was suggested she have her implantable neurostimulator (ins) checked by her healthcare provider (hcp).The patient stated that she is allergic to nickel but isn't sure why she would be getting this sensation now after everything was working so well for her.She just saw her hcp for an issue with her hip that she stated unrelated to the pain stim device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key10621312
MDR Text Key209626059
Report Number3004209178-2020-17146
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2021
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received09/30/2020
Date Device Manufactured04/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
-
-