• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/12HOLE/266MM/LT-STER; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/12HOLE/266MM/LT-STER; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.124.413S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter is company representative.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, during an unknown procedure, the threaded guide was not drilling to the guide.The surgeon used a different variable angle locking compression plate/12 hole/266 mm/left.The tip of the flexible shaft and stardrive screwdriver shaft was bent and the medullary reamer head was broke inside the patient.The fragments from the device were broken in the patient, surgery was delayed roughly about twenty to thirty (20-30) minutes.The procedure was successfully completed.This is report 1 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation flow: damage and device interaction/ functional.Visual inspection: the 4.5mm va-lcp curved condylar plate/12hole/266mm/lt-ster (part #: 02.124.413sand lot #: 2l51507) was received showing deformed threads in the first variable angle combi hole.No other issues were identified with the returned components of the device.The complaint condition could be confirmed.Functional test: a functional test cannot be performed as the 4.5mm va-lcp curved condylar plate/12hole/266mm/lt-ster was returned by itself.Can the complaint be replicated with the returned devices? unable to perform as the 4.5mm va-lcp curved condylar plate/12hole/266mm/lt-ster was returned by itself.However, the condition of the deformed thread could have caused the complaint condition.Dimensional inspection: dimensional inspection cannot be performed for threads due to the geometry of the device.Document/specification review: all relevant drawings are reflecting the manufactured and current revision, were reviewed.Complaint confirmed? yes.Investigation conclusion: the complaint condition is confirmed for the viper2 t20 hexlobe driver shaf (part #: 286725020 and lot #: ag2098246).There is no indication that a design or manufacturing issue has caused the issue and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot part number: 02.124.413s, lot number: 2l51507, manufacturing site: mezzovico, release to warehouse date: dec.14, 2018, expiry date: dec.01, 2028.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/12HOLE/266MM/LT-STER
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10621395
Report Number2939274-2020-04458
Device Sequence Number11414152
Product Code HRS
UDI-Device Identifier10886982042880
UDI-Public(01)10886982042880
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.124.413S
Device Catalogue Number02.124.413S
Device Lot Number2L51507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2020
Initial Date Received by Manufacturer 09/10/2020
Supplement Date Received by Manufacturer10/19/2020
Initial Report FDA Received Date10/02/2020
Supplement Report FDA Received Date10/21/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12.0MM MEDULLARY REAMER HEAD; 5.0MM FLEXIBLE SHAFT 620MM; STARDRIVE SCREWDRIVER SHAFT T8 55MM; UNK - GUIDE/COMPRESSION/K-WIRES; UNK - GUIDES/SLEEVES/AIMING; 12.0MM MEDULLARY REAMER HEAD; 5.0MM FLEXIBLE SHAFT 620MM; STARDRIVE SCREWDRIVER SHAFT T8 55MM; UNK - GUIDE/COMPRESSION/K-WIRES; UNK - GUIDES/SLEEVES/AIMING
-
-