| Model Number |
FX-410360 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "rupture".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "rupture".Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: underweight and opening extended on anterior.A visual and microscopic analysis was performed which identified: sharp opening on posterior assessed as a surgical impact opening, for the stress mark in the shell.A dimension measurement in the shell was performed which identified the thickness within specification.Based on the device analysis the final assessment is: a sharp opening on posterior assessed as a surgical impact opening.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "rupture".Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|