Model Number 283512
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Initial reporter phone number: (b)(6).Udi: (b)(4).
 
Event Description
It was reported by the customer as following: the gasket of the device came off after an unknown procedure.No harm to the patient reported.No delay reported.No further information available.Loaner requested.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the gasket of the device came off.Per service reports, this complaint can be confirmed.It was found during evaluation that the motor was corroded.Further, valve, head screw and levers were damaged.Cable and keyboard were in good condition.The motor, maintenance kit, valve and levers were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Fluid ingress into the device and contact with the motor is responsible for the rusty motor.The damaged valve, head screw and levers is most likely caused by external factors like user mishandling or a probable fall.A manufacturing record evaluation was performed for the finished device serial/lot number ((b)(6)), and no non-conformance was identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10623521
MDR Text Key209820255
Report Number1221934-2020-02868
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial,Followup,Followup
Report Date 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2020
Initial Date Manufacturer Received 09/18/2020
Initial Date FDA Received10/02/2020
Supplement Dates Manufacturer Received12/11/2020
02/25/2021
Supplement Dates FDA Received12/14/2020
02/26/2021
Removal/Correction NumberN/A
Patient Sequence Number1
-
-