• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC TRANSLACE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC TRANSLACE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 7480131
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient implanted with a set- screw using a driver for a spinal decompression and fusion (intra-op).It was reported that the provisional driver would not retain a set screw.There were no patient symptoms or complications as a result of the event.The pre-op diagnosis was spinal stenosis.The levels implanted were level 4/ 5.No further complications were reported/ anticipated.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient implanted with a set- screw using a driver for a spinal decompression and fusion (intra-op).It was reported that the provisional driver would not retain a set screw.There were no patient symptoms or complications as a result of the event.The pre-op diagnosis was spinal stenosis.The levels implanted were level 4/ 5.The product will be replaced by a medtronic product in the future.No further complications were reported/ anticipated.
 
Additional Manufacturer Narrative
H3: product analysis- visual and optical examination identified that the hex edges of the driver are worn from what appears to be repeated normal use.This is consistent with anticipated wear.The center jaw/prongs of the driver have been compressed and bent not allowing it to retain the screw anymore.This appears to be the result of repeated use overtime.This is consistent with anticipated wear.H6: analysis resulted in change in eval.Code method, eval.Code result, patient code medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSLACE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10625299
Report Number1030489-2020-01367
Device Sequence Number16956074
Product Code LXH
UDI-Device Identifier00721902853579
UDI-Public00721902853579
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7480131
Device Catalogue Number7480131
Device Lot NumberM04K0346
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2020
Supplement Date Received by Manufacturer12/18/2020
Initial Report FDA Received Date10/05/2020
Supplement Report FDA Received Date12/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-