Catalog Number 3505-6545
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports: 3012447612-2020-00584 and 3012447612-2020-00585.
 
Event or Problem Description
It was reported three pathfinder next screws fractured during surgery.One of the screws broke in the surgeon¿s hands while assembling.The other two screws broke in-situ during the tightening of the nut.The fractured screws were removed, and replaced with new ones.There were no reported patient impacts.This is report one of three for this event.
 
Additional Manufacturer Narrative
Information was entered erroneously into h3; there are no changes from the initial report.Additional information in h6: method, results, and conclusions.This follow-up report is being submitted to relay additional information and initially corrected information.Device evaluation:the product number and lot numbers in the photograph match the complaint.Visual inspection reveals that the tulip head was detached from the screw.The complaint is confirmed for all three products based on the available pictures.Complaint history search: the complaint history search for disassembly of tulip head was performed.Dhr review and related actions per dhr review, the parts were likely conforming when it left zimmer biomet control.No actions required.This event is not related to any current actions or recalls or product holds.Potential cause with the given information, the definitive cause cannot be determined.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Event or Problem Description
It was reported three pathfinder nxt screws fractured during surgery.One of the screws broke in the surgeon's hands while assembling.The other two screws broke in-situ during the tightening of the nut.The fractured screws were removed and replaced with new ones.There were no reported patient impacts.This is report one of three for this event.
 
Event or Problem Description
It was reported three pathfinder nxt screws fractured during surgery.One of the screws broke in the surgeon¿s hands while assembling.The other two screws broke in-situ during the tightening of the nut.The fractured screws were removed and replaced with new ones.There were no reported patient impacts.This is report one of three for this event.
 
Event or Problem Description
It was reported three pathfinder nxt screws fractured during surgery.One of the screws broke in the surgeon's hands while assembling.The other two screws broke in-situ during the tightening of the nut.The fractured screws were removed and replaced with new ones.There were no reported patient impacts.This is report one of three for this event.
 
Additional Manufacturer Narrative
Device evaluation: visual inspection reveals that the tulip head was detached from the screw.Device use: this devices are used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED POLYAXIAL SCREW 6.5MM X 45MM
Common Device Name
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
jessica buzbee
10225 westmoor dr.
na
westminster, CO 80021
3038034523
MDR Report Key10626613
Report Number3012447612-2020-00583
Device Sequence Number2604201
Product Code NKB
UDI-Device Identifier00889024335707
UDI-Public(01)00889024335707
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K132884
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3505-6545
Device Lot NumberAAJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/25/2020
Supplement Date Received by Manufacturer10/05/2020
12/11/2020
04/27/2022
Initial Report FDA Received Date10/05/2020
Supplement Report FDA Received Date10/15/2020
12/17/2020
05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-