|
|
| Model Number |
97725 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); High impedance (1291)
|
| Health Effect - Clinical Codes |
Therapeutic Effects, Unexpected (2099); Insufficient Information (4580)
|
| Date of Event |
09/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: 977d260; serial#: (b)(4); iimplanted: (b)(6) 2020; explanted: (b)(6) 2020; product type: screening device.Product id: 977d260; serial# (b)(4); implanted: (b)(6) 2020; explanted: (b)(6) 2020; product type: screening device.Product id: 977d260; serial#: (b)(4); implanted: (b)(6) 2020; explanted: (b)(6) 2020.Product type: screening device.Other relevant device(s) are: product id: 977d260; serial/lot #: (b)(4); ubd: 22-sep-2024, udi#: (b)(4); product id: 977d260; serial/lot #: (b)(4); ubd: 10-sep-2024; udi#: (b)(4); product id: 977d260; serial/lot #: (b)(4); ubd: 10-sep-2024; udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The manufacturing representative (rep) reported a spinal cord stimulation trial was performed on (b)(6) 2020.When one lead was placed in the epidural space, and connected to a wireless external neurostimulator (wens), stimulation was activated.The patient could feel no paresthesia.Impedances were checked, and were found to be all >10,000.Lead connectivity was good.First lead was explanted, and second lead was placed with the same results.An additional wens was tried, which produced the same results.The second lead was explanted, and a third lead was placed producing the same results.The trial was aborted, and the issue resolved.Physician requested that the product be returned for analysis.There were high impedances, and no stimulation felt.There were no known environmental, or patient factors led to, or contributed to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of product id 977d260 serial# (b)(6), product id 977d260 serial# (b)(6), product id 977d260 lot# serial# (b)(6) found no anomaly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|