• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 500 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 500 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938738
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 09/09/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter city: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 500ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was leaking from an unspecified location.The leak was identified during setup/preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d10, h3, h4 and h6.H4: device manufactured between february 11, 2020 to february 12, 2020.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak at the spike port bonding area.The reported condition was verified.The cause was not determined; however the most likely cause was due to inadequate or lack of cyclohexanone being applied to the cap tubing when it was inserted to the spike port during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
500 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10627403
MDR Text Key209803940
Report Number1416980-2020-06175
Device Sequence Number1
Product Code LHI
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2023
Device Catalogue NumberH938738
Device Lot Number60226707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2020
Date Manufacturer Received10/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-