BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/26; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364505
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
The orsiro drug-eluting stent system was chosen for the treatment.During lesion crossing the stent became deformed at the proximal end.
 
Additional Manufacturer Narrative
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the stent is severely deformed at its proximal end.Several stent struts are lifted.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the lifted struts were initially crimped on the balloon.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.25/26
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10627961
Report Number1028232-2020-04263
Device Sequence Number3297316
Product Code NIQ
UDI-Device Identifier07640130439146
UDI-Public7640130439146
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/11/2020
Device Model Number364505
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09184681
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2020
Initial Date Received by Manufacturer 10/05/2020
Supplement Date Received by Manufacturer12/04/2020
Initial Report FDA Received Date10/05/2020
Supplement Report FDA Received Date12/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-