• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 5/80/90; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 5/80/90; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 357473
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2020
Type of Reportable Event Malfunction
Event or Problem Description
A passeo-18 balloon catheter was chosen for the treatment of a severely calcified lesion in the superficial fermoral artery.During the attempt of postdilatation of a stent, the passeo-18 balloon ruptured at 6 atm without causing any harm to the patient.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 41 mm.Microscopic inspection of the balloon showed several deep scratches in close vicinity of the tear.It seems likely that the tear in the balloon was caused by a hard, sharp edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause is most likely related to the patients anatomy (i.E.Calcified lesion).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 5/80/90
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10627964
Report Number1028232-2020-04264
Device Sequence Number3297317
Product Code LIT
UDI-Device Identifier07640119558844
UDI-Public07640119558844
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model Number357473
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09185092
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2020
Initial Date Received by Manufacturer 10/02/2020
Supplement Date Received by Manufacturer11/16/2020
Initial Report FDA Received Date10/05/2020
Supplement Report FDA Received Date11/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-