| Model Number |
10606 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
09/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that stent fracture occurred.The target lesion was located in the moderately calcified left main into the left anterior descending artery.A 3.00 x 32 synergy drug eluting stent was advanced for treatment and successfully deployed.However, after intravascular ultrasound was used together with an enhance stent visualization tool, it was noted that the stent was fractured.Subsequently, a second layer of stent was implanted to cover the fracture and the procedure was completed.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that stent fracture occurred.The target lesion was located in the moderately calcified left main into the left anterior descending artery.A 3.00 x 32 synergy drug eluting stent was advanced for treatment and successfully deployed.However, after intravascular ultrasound was used together with an enhance stent visualization tool, it was noted that the stent was fractured.Subsequently, a second layer of stent was implanted to cover the fracture and the procedure was completed.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the media provided by the site could not identify the alleged stent fracture.A gap which could indicate strut displacement was noted in the proximal to mid stent region and this might have appeared to the physician as stent fracture.The displaced struts visible in the photo review most likely represent a conformation of the stent to the fibro-calcific lesions in the distal section of the vessel anatomy allowed by the stent 4-connector proximal and 2-connector throughout.The coplanar view of the stent could lead to the assumption that there was a stent fracture when in fact what was being seen was two strut connectors on the same plane on view; this would be consistent with the stent image reviewed where struts are displaced in a v shape and connected in a central point.
|
| |
|
Search Alerts/Recalls
|