It was reported that during treatment with a prismaflex m100 set c, an external blood leak was observed from a ¿gap at the tube link above the filter." there was no patient injury, or medical intervention associated with this event.No additional information is available.
|
H10: the actual device was not available; however, a picture and video of the sample were provided for evaluation.Visual inspection of the provided picture showed that there was an external blood leak at the level of the return screwed connector.Based on the video it seemed the connector was well screwed, and not damaged.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental re.
|