• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955595
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/09/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility name: (b)(6).Initial reporter phone number: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with a prismaflex m100 set c, an external blood leak was observed from a ¿gap at the tube link above the filter." there was no patient injury, or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, a picture and video of the sample were provided for evaluation.Visual inspection of the provided picture showed that there was an external blood leak at the level of the return screwed connector.Based on the video it seemed the connector was well screwed, and not damaged.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental re.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10628818
MDR Text Key209840627
Report Number8010182-2020-00200
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955595
Device Lot Number19A1409
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/10/2020
Initial Date FDA Received10/05/2020
Supplement Dates Manufacturer Received10/22/2020
Supplement Dates FDA Received10/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-