ZHONGSHAN A&J MEDICAL EQUIPMENT DRIVE; ROLLATOR Back to Search Results
Model Number RTL10266CF
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Codes Bruise/Contusion (1754); Concussion (2192)
Date of Event 01/19/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Drive devilbiss healthcare is the initial importer of the device which is a rollator.This report is being filed as part of an mdr regression analysis.End-user was walking her dog with the device.She reported that she was leaning on the left hand side of the device and the frame near the handle bars gave way.She fell to the ground, was unconscious, and broke her nose.She was taken to the hospital by ambulance.She was diagnosed with a concussion and a bruised nose.The unit was returned and evaluated and classified as defective : defective: a product that after use shows an inability to meet its intended design functionality due to a determined factor.The break was on the left frame not the handle.There was a confirmed defect on the left side of the frame.However, the defect did not go through to the other side of the frame nor did it cause the frame to fall apart.This may have been due to a horizontal force placed on the unit, causing stress and the unit to crack.There is a baggage claim tag on the unit for flight dl 0574 as well as a wire on the backrest adjustment knob.Historical data on complaint shows that this was an isolated incident.Root cause cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVE
Common Device Name
ROLLATOR
Manufacturer (Section D)
ZHONGSHAN A&J MEDICAL EQUIPMENT
3 shenghui south rd
nantoou town
zhongshan city, guangdong
CH 
MDR Report Key10630658
Report Number2438477-2018-00071
Device Sequence Number3649384
Product Code ITJ
UDI-Device Identifier00822383576817
UDI-Public822383576817
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/05/2020
Report Date (Section F) 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberRTL10266CF
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/05/2018
Device Age14 MO
Event Location Other
Date Report to Manufacturer10/06/2020
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
Patient Weight90
Date Report Sent to FDA10/05/2020
-
-