MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; ELEOS SEGMENTAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; ELEOS SEGMENTAL STEM Back to Search Results
Model Number KSC18100E
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The root cause for the dissociation of the resurfacing femur and the stem extension was unable to be determined.The patient has undergone two previous revision surgeries for dissociation since her primary surgery all performed by dr.Siegel.The patient's medical history is unknown, and it is unknown whether the patient sustained a trauma prior to the failure.Based on review of the device history records, and sterilization batch release records, it was concluded that the failure was not associated with the manufacture of the implants, or a nonconformance.
 
Event or Problem Description
On (b)(6) 2020, a (b)(6)-year-old female underwent a revision surgery due to dissociation of stem extension and resurfacing femur dissociated.During the revision surgery, the tibial hinge component, tibial poly spacer, resurfacing femur, and resurfacing femur axial pin were all replaced with identical implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
ELEOS SEGMENTAL STEM
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
MDR Report Key10631451
Report Number3013450937-2020-00135
Device Sequence Number3529454
Product Code KRO
UDI-Device IdentifierB278KSC18100E0
UDI-PublicB278KSC18100E0
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2026
Device Model NumberKSC18100E
Device Catalogue NumberKSC18100E
Device Lot Number1750157
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
-
-