• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC ROD CUTTER; CUTTER, WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC ROD CUTTER; CUTTER, WIRE Back to Search Results
Model Number 388.72
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Part: 388.72.Lot: t186215.Manufacturing site: (b)(6).Release to warehouse date: 09-mar-2020 (5 devices), 30-mar-2020 (10 devices) and 15-apr-2020 (5 devices) a review of the device history records was performed for the finished device lot number and no relevant nonconformances were detected.Visual inspection: the rod cutter was returned and received at us customer quality (cq).Upon visual inspection, it was observed that the distal tip of the device was broken and the broken fragment was not returned.No other issues were identified with the returned device.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage.Document/specification review based on the date of manufacture, the current and manufactured revision of drawings were reviewed.Rod cutter.Investigation conclusion: the complaint condition was confirmed for the rod cutter.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.The potential cause could be due to unintended forces applied to the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from (b)(4) reports an event as follows: it was reported that during a procedure on (b)(6) 2020, the surgeon had to cut a stainless steel needle and damaged the distal tip of the rod cutter blades.The rod cutter had been recently purchased and was damaged the first time it was used.Procedure was completed successfully with no delay.There was no impact on the patient.Upon manufacturer investigation, it was determined that the device was broken.This report is for a rod cutter.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD CUTTER
Common Device Name
CUTTER, WIRE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10631632
Report Number2939274-2020-04499
Device Sequence Number3205443
Product Code HXZ
UDI-Device Identifier10705034773759
UDI-Public(01)10705034773759
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number388.72
Device Catalogue Number388.72
Device Lot NumberT186215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/05/2020
Initial Report FDA Received Date10/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-