• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) Back to Search Results
Model Number DM0010FAA
Medical Device Problem Code Device Remains Activated (1525)
Health Effect - Clinical Codes Perforation (2001); No Consequences Or Impact To Patient (2199)
Date of Event 09/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was not returned.This dm0010faa easydrill cranial perforator with lot number 1674/19 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the perforator failed to auto-stop and plunged into the brain tissue.It was also mentioned that the perforator was used with a stryker pi drill and an attachment.
 
Event or Problem Description
On follow up, it was reported that the event happened during a craniotomy procedure and there was no delay or additional interventions performed.It was also reported that the device did not hit the patient's dura and no further information can be provided regarding the status of the patient post-surgery.
 
Additional Manufacturer Narrative
Report not confirmed.Evaluation could not reproduce the reported malfunction of continues to run.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR  09980-39
MDR Report Key10631844
Report Number1625507-2020-00183
Device Sequence Number3297434
Product Code HBF
UDI-Device Identifier07898959543333
UDI-Public07898959543333
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K141455
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/29/2021
Device Model NumberDM0010FAA
Device Catalogue NumberDM0010FAA
Device Lot Number1674/19
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2020
Supplement Date Received by Manufacturer10/07/2020
Initial Report FDA Received Date10/05/2020
Supplement Report FDA Received Date11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STRYKER DRILL (S/N: UNKNOWN) ATTACHMENT (UNKNOWN)
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient Weight90
-
-