MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3544751 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: migration of breast prostheses.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old caucasian female patient underwent a breast augmentation revision procedure with a mentor memorygel breast implant 475cc gel breast prosthesis (left device) and a mentor memorygel breast implant 375cc gel breast prosthesis (right device) that both dropped lower in the breast pocket after implantation.Bilateral breast prosthesis migration was diagnosed by a physical exam.As a result, the patient underwent bilateral explantation and replacement with a mentor memorygel breast implant 475cc gel breast prosthesis and a mentor memorygel breast implant 375cc gel breast prosthesis on (b)(6) 2020.This medwatch report is for the patient¿s left breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On 14-oct-2020, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information reported, the patient's implants dropped lower in the breast pocket.During the visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Implant displacement/migration is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information is consistently analyzed and monitored by quality assurance to determine when further action is necessary.Manufacturer¿s reference number: (b)(4).
|
| |
|