Model Number DM3500 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the implantable cardiac monitor exhibited wireless communication issues with the mobile app.The patient had received a transmission error daily.It was unknown whether the issue was due to the device or the app.The patient was provided with a new remote transmitter.The patient was discontent but asymptomatic.
|
|
Event Description
|
New information received notes that the issue resolved when the operating system was updated from 13 to 14 on an iphone 12 after trying many iphone models.The physician believes the app is the issue.
|
|