• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37612
Device Problems Failure to Charge (1085); Failure to Deliver Energy (1211); Charging Problem (2892)
Patient Problem Ambulation Difficulties (2544)
Event Date 09/15/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Recharger defect (error code 376) was reported.It was stated that charging became unable to be performed, and the device turned off, patient could not move, so the patient was transferred to hospital by ambulance for treatment.When the reset button on the charger was pressed down, it was able to charge normally once, but error code 376 occurred.The product was replaced with a new product and it was confirmed that the new product could be charged without any problems.Issue was resolved.
 
Manufacturer Narrative
Section d information references the main component of the system, other applicable components are: product id 37651, lot# serial# (b)(6), implanted: n/a, explanted: n/a, product type recharger d9: the recharger serial # (b)(6) was returned.The ins remains in use.H3: analysis of the recharger identified a failure of the recharger antenna.Analysis determined there was no communication with a known good test ins.H6: eval codes b01, c07, d16 apply to the recharger, and b20, c20, d14 to the implantable neurostimulator (ins).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H6: evaluation code conclusion updated per additional investigation activities.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10634765
MDR Text Key209987688
Report Number3004209178-2020-17299
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2017
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2020
Date Manufacturer Received01/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age50 YR
-
-