Device evaluated by mfr: the device was returned for analysis.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.The device was soaked in a water bath for 14 days to loosen the blood and contrast in the device.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There were no issues detected.The device was x-rayed, and the marker bands were detected.Inspection of the remainder of the device presented no other damage or irregularities.
|